To access the Time and Date function, press SET while the device is ON. By pressing the SET button again, you can access specific functions whose parameters can be adjusted pressing and .
E.g.: If you only want to set the time, press SET four times and adjust the value using and .FREQUENCY | TYPE OF INTERVENTION | INTERVENTION PROCEDURE |
---|---|---|
Every day | Manual cleaning of the sterilization chamber. | Cleaning operation to be performed manually when the chamber is cold using a cloth dampened with demineralised water |
Every day | Manual cleaning of the door’s rubber seal. | Cleaning operation to be performed manually when the chamber is cold using a cloth dampened with demineralised water |
Every week | Cleaning or replacement of the demineralised water filter | See Paragraph 7.3 of the user manual |
When the NEED CLEANING message appears on the display. | Clean the sterilization chamber using the automatic periodic cleaning and Demineralised water filter cleaning cycles to be performed when the chamber is cold | See Paragraph 7.2 of the user manual |
Every 6 months or after 500 cycles | Replace the bacteriological filter | See Paragraph 7.4 of the user manual |
We recommend referring to the user manual for further details
To view the total number of cycles performed by the sterilizer since it was installed, use the following button sequence making sure the device is ON:
The purpose of the Helix Test and Bowie&Dick Test is to verify the perfect penetration of the sterilizing steam into the hollow parts of the tools (Helix Test) and into the porous material (Bowie&Dick Test) in the sterilization chamber.
The Helix Test and Bowie&Dick Test can be activated when the sterilizer is on. We recommend performing them at the start of the day, before starting the daily sterilization cycles, anyway they can be carried out at any moment when the sterilizer is on.
Prepare the sterilizer for a Helix Test or Bowie&Dick Test according to what requested by standards. While the machine is on, press TEST and START/STOP.
The device will automatically initiate the test programme, characterized by a temperature of 134°C, a sterilization time of 3.5 minutes and 3 vacuum phases.
If you have started a Helix Test or Bowie&Dick Test by mistake, you can interrupt it by pressing START/STOP. The device eliminates the test started by mistake, automatically starts an evacuation procedure of the liquids in circulation and places itself on stand-by, ready to be used for a normal work cycle
In order to preserve the properties of the equipment and avoid problems, we recommend reading the following indications carefully:
First of all, check that DXP NewSeal is connected to the mains.
Then check that the fuses are not damaged.
If the problem persists, contact the technical assistance.