ps-bg-yellow

Demin30

New demineralization system

Demin30 is a demineralization system with mixed-bed and ion-exchange resin, which enables the complete demineralization of the water. Easy to install on various autoclave models, offering optimal performance and protection of sterilised instruments, extending the life of autoclaves or other reprocessing devices. The filter can be positioned on the worktop or inside the sterilisation cabinet. Demin30 can also supply demineralized water to a second autoclave.

PLUS

Efficiency
Demin30 filters up to 580 litres* of water with a single cartridge, guaranteeing an excellent level of purification. You can save time thanks to the automatic loading and unloading of the water, maximising efficiency and convenience.

*Calculated based on water with a conductivity of 250μS/cm
Versatility
Modular system for supplyng one or two autoclaves, also available in an EN1717 compliant version for water back-flow prevention. Installation adaptable to various autoclave models. The filter can be positioned on the worktop or inside the sterilisation cabinet.
Usability
Cartridges can be replaced quickly with no need for tools. The cartridge must be replaced when the water conductivity exceeds 15 μS/cm. DXPclave gives an automatic warning when this value is reached to indicate that the filter needs to be replaced.

AVAILABLE KITS

MODEL Demin30 – Kit EN1717

CODE 98001010


MODEL Demin30 – Kit 2nd Autoclave EN1717

CODE 98001030

MODEL Demin30 – Kit

CODE 98001000


MODEL Demin30 – Kit 2nd Autoclave

CODE 98001020

MODEL Demin30 – Demineralization Resin Filter

CODE 230000

MAIN COMPLIANCES

The kit supplied with a top-loading water system complies with EN1717 Standard: “Protection against pollution of potable water in water installations and general requirements for devices to prevent pollution caused by back-flow”

HOW TO INSTALL IT

The filter must only be supplied with potable water, with an inlet pressure between 2 and 3 bar.

Thanks to the supplied hose, it is possible to drain used water automatically. The solenoid valve in the kits is connected directly to the autoclave and does not require an external power supply.

The use of Demin30 is recommended only in the presence of water with a conductivity <600μS/cm.

The 2nd autoclave kit can be used either to connect a second autoclave to Demin30 or to enable automatic loading of an autoclave if there is a centralised demineralization or osmosis system within the clinic.

AUTONOMY

Autonomy**

(based on the conductivity of the incoming water)

200μS/cm720 L
250μS/cm580 L
300μS/cm480 L
400μS/cm360 L
500μS/cm290 L
600μS/cm240 L

Request information